SCPI Command Reference

194

IQS-1600

Product-Specific Commands—Description

:SENSe[1..n]:POWer:WAVelength

Description

This command selects a new operating

wavelength on the specified channel. If no

channel was specified, the default value used is

1.

Syntax

:SENSe[1..n]:POWer:WAVelength

length>|MAXimum|MINimum|DEFault

Parameter(s)

Wavelength:
The program data syntax for is

defined as a element. The

special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the parameter.

The parameter is an operating
wavelength using meter as units. Any

wavelength within the spectral range of the

instrument optical detector at a resolution of 0.01

nm may be selected.

Example(s)

SENS:POW:WAV 0.00000131002

or

SENS:POW:WAV 1310.02 nm

Notes

See the instrument's user guide for the exact

spectral range for each detector type.

See Also

SENSe:POWer:WAVelength?