Creating and Starting a Test Case

80

FTB-8500 Series and FTB-8120NGE/8130NGE

Ethernet EtherSAM (Y.1564) and RFC 2544 Dual Test Set Test Cases

2. Port configuration:

Port 1 (FTB-8510B only) is automatically selected.

2a. Select the port Connector

Type (Optical or
Electrical RJ-45). Optical is automatically selected for the
Ethernet 10G interface.

2b. For Electrical RJ-45, if required, select

the Ethernet port crossover check box
to inverse the pin-to-pair assignment of
the UTP cable used.

Note: At this point you should have a link up indicated in the Tree view tab. A

green LED indicates a link up while a red LED indicates a link down. For
electrical port, if the link is down, make sure that the Ethernet port
crossover
setting is correct.

2c. Press Next.