Manual power failure recovery, When using the test timer, When using smartmode – EXFO FTB-8500 Series for FTB-500 User Manual
Page 456
Automatic Power Failure Recovery
442
FTB-8500 Series and FTB-8120NGE/8130NGE
Manual Power Failure Recovery
If only the requirement number 1 is met, the saved configuration will be
loaded when the FTB-8500 Series and FTB-8120NGE/8130NGE is manually
started from ToolBox. Thus, the test that was running will be re-created,
configured, and restarted.
Note: For RFC 2544, the configuration will be re-loaded but the test will not
restart automatically.
Note: The power failure recovery is disabled when the SUI terminates normally
or when the test case is cleared.
When Using the Test Timer
Refer to Timer Configuration on page 124 for more information on test
timer.
The test that was running will be re-created and started after a power
failure if all the following conditions are met:
The test was running.
Startup Application is enabled on the FTB-500.
The start time has not expired during the power failure.
The stop time or the duration has not expired during the power failure.
When Using SmartMode
SmartMode is not supported meaning that SmartMode will return to its
default factory settings after a power failure recovery.