Throughput (rfc 2544) – EXFO FTB-8500 Series for FTB-500 User Manual
Page 440
Expert Mode Tabs
426
FTB-8500 Series and FTB-8120NGE/8130NGE
Throughput (RFC 2544)
Throughput (RFC 2544)
Press TEST, Expert Mode, and Throughput.
Enable Criteria: Allows to enable Throughput Pass/Fail criteria. This
setting is disabled by default.
Note: Both Enable Criteria and the test itself (see Global Configuration on
page 298) have to be enabled for a specific test to be able to change its
criteria values.
Frame Size: Indicates the frame sizes selected for the test.
TX-to-RX/P1-to-P2/P2-to-P1/Local to Remote/Remote to Local: For
each frame size, enter the minimum throughput values allowed before
declaring a Fail condition. The default values are the maximum
throughput. P1-to-P2/P2-to-P1 is only available with FTB-8510B.
Status: Indicates the real-time pass/fail condition.
Grey LED: Indicates that Expert Mode is disabled or the test did not
run yet.
Green LED: Indicates that the threshold is respected (Pass).
Red LED: Indicates that the threshold is not respected (Fail).
Copy From Measurement button: Copy the performance results from
the test that ran previously and use it as the Pass/Fail criteria. If no test
ran before, default values used will be 0 bps.