Logger table – EXFO FTB-8500 Series for FTB-500 User Manual
Page 144
Summary Tabs
130
FTB-8500 Series and FTB-8120NGE/8130NGE
Test Logger
Logger Table
An event is automatically listed in the logger and saved on the hard drive in
case a power failure condition occurs.
The logger is cleared when one of the following conditions is met:
A test case is stopped and restarted.
The test case is cleared.
When pressing
.
Events are listed by ID - Date/Time by default. Events can also be sorted by
Data Path or Event by pressing on the corresponding column title.
ID: Indicates the Event number. Events are sequentially numbered.
Date/Time: Indicates the date and time the Alarm/Error condition has
been detected.
Data Path: Indicates the origin of the alarm/error. [P1] and [P2] in the
data path represent respectively the Port 1 and Port 2. Port 2 is only
available with FTB-8510B.
Event: Indicates the alarm/error type.
Duration: Indicates the number of seconds (day:hour:minute:second
format) within which the alarm/error occurred.
Count: Indicates the number of occurrences of the error.
Rate: Indicates the error rate.
Note: In the Duration, Count and Rate columns, Pending indicates that the
alarm/error condition persists or was persisting when the test was stopped.