SCPI Command Reference

616

FTB-5240S/S-P/BP

Examples on Using the SCPI Commands

Performing a Continuous Acquisition with
Synchronized Intermediate Results Query (WDM)

Command Sequence

Comments

analysis parameters>

See Configuring the Analysis Setup for the Next
Acquisition Sequence (WDM)
on page 603.

TRAC:FEED:CONT "TRC1", NEXT

Disable continuous refresh of WDM analysis

active trace; set feed control for "one-shot"

refresh.

SENS:CORR:OFFS:MAGN 0.0 DB

Disable power offset.

SENS:WAV:OFFS 0.0 NM

Disable wavelength offset.

SENS:FREQ:STAR 190.9506 THZ

Set sweep frequency range.

SENS:FREQ:STOP 196.5852 THZ

""

SENS:AVER:STAT OFF

Enable trace averaging.

TRIG:SEQ:SOUR IMM

Set sweep trigger event source to immediate.

STAT?

Test instrument state is idle. Poll STAT? until the

returned state is READY.

INIT:CONT ON

Start sweep acquisition loop.

TRAC:FEED:CONT? "TRC1"

Wait for the first trace refresh to be done. Poll

WDM analysis trace feed until the returned

value is NEVER.

results>

See Retrieving Analysis Results (WDM) on

page 609.

TRAC:FEED:CONT "TRC1", NEXT

Reactivate WDM analysis trace feed control for

another "one-shot" refresh.

TRAC:FEED:CONT? "TRC1"

Wait for trace refresh done. Poll trace feed until

the returned value is NEVER.

results>

See Retrieving Analysis Results (WDM) on

page 609.