beautypg.com

EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 441

background image

SCPI Command Reference

OSA

435

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:WIDTh:FREQuency?

Description

This query returns the frequency width of the

noise measurement region of the selected WDM

analysis channel.

At *RST, this value is not available.

At CALCulate[1..n][:WDM]:CHANnel[:DEFine],

width of the noise measurement region is set to

100.0 GHz.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT

h:FREQuency?[MAXimum|MINimum|DE

Fault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax