Initiate:continuous – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 530

background image

SCPI Command Reference

524

FTB-5240S/S-P/BP

Product-Specific Commands—Description

SENS:WAV:STAR 1525.000 NM

SENS:WAV:STOP 1570.000 NM

TRIG:SEQ:SOUR IMM

STAT? Poll until returned state is READY

INIT:CONT ON

INIT:CONT? Returns 1 (trigger system

continuously initiated)

...

INIT:CONT OFF

STAT:OPER:BIT8:COND? Poll until returned state

is 0

Notes

The trigger system leaves IDLE state to perform

acquisition only if the instrument is in READY

status.

Trace averaging is not supported by the trigger

system when continuously initiated.

Continuous acquisition does not support InBand

noise analysis: acquired trace is always analysed

using IEC noise measurement.

See Also

:ABORt

:CALCulate[1..n][:WDM]:STATe

:INITiate[:IMMediate]

:INITiate:CONTinuous?

:SENSe[1..n]:AVERage:STATe

:STATus?

:STATus:OPERation:BIT<8|9>:CONDition?

:TRACe:FEED:CONTrol

:TRIGger[1..n][:SEQuence]:SOURce

:INITiate:CONTinuous