beautypg.com

EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 611

background image

SCPI Command Reference

OSA

605

Examples on Using the SCPI Commands

Creating a Channel List Based on Specific Channels
(WDM)

Command Sequence

Comments

CALC:WDM:CHAN:AUTO OFF

Disable default channel.

CALC:WDM:CHAN:DEL:ALL

Clear current channel list.

CALC:WDM:CHAN:DEF
"CWDM_1470",1470.0 NM

Add a new channel named "CWDM_1470" with

nominal central wavelength at 1470.0 nm.

All others parameters for this new channel are

set to their default values.

CALC:WDM:CHAN:SEL "CWDM_1470" Select channel "CWDM_1470".

parameters>

See Modifying Selected Channel Analysis
Parameters (WDM)
on page 608.

CALC:WDM:CHAN:DEF
"CWDM_1530",1530.0 NM

Add a new channel named "CWDM_1530" with

a nominal central wavelength at 1530.0 nm.

All others parameters for this new channel are

set to their default value.

CALC:WDM:CHAN:SEL "CWDM_1530" Select channel "CWDM_1530".

parameters>

See Modifying Selected Channel Analysis
Parameters (WDM)
on page 608.

CALC:WDM:CHAN:DEF
"CWDM_1550",1550.0 NM

Add a new channel named "CWDM_1550" with

a nominal central wavelength at 1550.0 nm.

All others parameters for this new channel are

set to their default value.

CALC:WDM:CHAN:SEL "CWDM_1550" Select channel "CWDM_1550".

parameters>

See Modifying Selected Channel Analysis
Parameters (WDM)
on page 608.