Sense[1..n]:frequency:stop – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 561

background image

SCPI Command Reference

OSA

555

Product-Specific Commands—Description

:SENSe[1..n]:FREQuency:STOP

Description

This command sets instrument sweep stop

frequency.

At *RST, this value is set to 196.5852 THz.

Syntax

:SENSe[1..n]:FREQuency:STOP

wsp>HZ]>|MAXimum|MINimum|DEFault

Parameter(s)

Stop:
The program data syntax for is defined

as a element followed by an

optional element.

The allowed

element is HZ. The special forms

MINimum, MAXimum and DEFault are accepted

on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.

DEFault allows the instrument to select a value

for the parameter.

The parameter corresponds to a valid

sweep stop frequency value.

The SENSe[1..n]:FREQuency:STOP? MIN and

SENSe[1..n]:FREQuency:STOP? MAX queries can

be used to determine valid sweep stop

frequency range.