Setting Up the Instrument in Drift Mode

OSA

131

Setting Up Acquisition Parameters

Setting Up Acquisition Parameters

Before performing your test, you must set the acquisition type and
parameters from the Acquisition tab and other parameters from the Drift
Settings
tab.

There are three types of acquisitions in Drift mode: single, averaging and
InBand.

Single: Spectral measurement is performed once. The results are
displayed according to this measurement.

Averaging: Spectral measurements are performed based on the
number of scans that you have entered for this parameter. The trace
will be displayed after each acquisition and averaged with the previous
traces.

InBand: The InBand type acquisition will perform a series of scans in
different polarization conditions in order to enable the InBand OSNR
calculation.

Note: The InBand option is available only if the module supports it.

Before performing measurements on an optical spectrum, you must select
the wavelength/frequency range to use. You can perform the scan on the
full range, on spectral bands, or select a custom range.

Note: The shorter the wavelength or frequency range, the faster the acquisition.

You can configure the delay, sampling and total duration for a drift
measurement. You can also configure the drift files name and select a
location where it should be saved.

The application allows you to perform a trial scan while setting up the drift
measurement.