beautypg.com

Building a custom drift measurement – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 144

background image

Setting Up the Instrument in Drift Mode

138

FTB-5240S/S-P/BP

Building a Custom Drift Measurement

Building a Custom Drift Measurement

You can build a drift measurement using a WDM measurement you already
have as a reference. The selected channels and thresholds can be
imported from the analysis setup or the reference measurement.

A custom drift measurement is particularly useful for offline processing of
your data over time and comparing result variations.

The WDM measurements you add must meet specific criteria to be
included in the custom build. The table below describes those
compatibility criteria.

Note: Files that are not compatible will automatically be rejected from the custom

build measurement.

Criteria

Test

Compatibility

Status

Acquisition type

Target WDM measurement

acquisition type differs from drift

reference trace acquisition type

Compatible

with warnings

Acquisition number

of scans

Target WDM measurement

acquisition number of scans differs

from drift reference trace

acquisition number of scans.

Compatible

with warnings

Spectral range

Target WDM measurement

acquisition spectral range only

partially overlaps spectral range

of drift reference trace.

There is no overlap between

target WDM measurement

spectral range and drift

reference trace spectral range.

Compatible

with

warnings

Incompatible