:CALCulate[1..n][:WDM]:THReshold

Description

This command sets WDM analysis absolute

power threshold for peak detection.

At *RST, this value is set to -45.0 dBm.

Syntax

:CALCulate[1..n][:WDM]:THReshold

reshold[DBM|W]>|MAXimum|MINimu

m|DEFault

Parameter(s)

Threshold:
The program data syntax for is

defined as a element

followed by an optional

DATA> element. The allowed

PROGRAM DATA> elements are: DBM|W. The

special forms MINimum,

MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the

smallest supported value.

MAXimum allows to set the instrument to the

greatest supported value.