Calculate[1..n][:wdm]:threshold – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual
Page 518

SCPI Command Reference
512
FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:THReshold
Description
This command sets WDM analysis absolute
power threshold for peak detection.
At *RST, this value is set to -45.0 dBm.
Syntax
:CALCulate[1..n][:WDM]:THReshold reshold[ m|DEFault Parameter(s) Threshold: defined as a followed by an optional DATA> element. The allowed PROGRAM DATA> elements are: DBM|W. The MAXimum and DEFault are accepted on input. MINimum allows to set the instrument to the smallest supported value. MAXimum allows to set the instrument to the greatest supported value.
The program data syntax for