SCPI Command Reference

OSA

569

Product-Specific Commands—Description

:SENSe[1..n][:WAVelength]:STOP?

Description

This query returns a value indicating either the

current or the minimum/maximum instrument

sweep stop wavelength.

At *RST, this value is set to 1570.0 nm.

Syntax

:SENSe[1..n][:WAVelength]:STOP?[MAXi

mum|MINimum|DEFault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's

smallest supported value.

MAXimum is used to retrieve the instrument's

greatest supported value.

DEFault is used to retrieve the instrument's

default value.

Response Syntax