FTB-5240S/S-P/BP
Examples on Using the SCPI Commands
Performing a Custom InBand Acquisition (WDM)
STAT:OPER:BIT8:COND?
Wait for the acquisition to be completed. Poll
bit 8 until the returned value is 0.
results> See Retrieving Analysis Results (WDM) on Command Sequence Comments analysis parameters> See Configuring the Analysis Setup for the Next CALC:WDM:CHAN:AUTO:NOIS:AUTO Optional: if the default channel is active then set auto noise to disabled. CALC:WDM:CHAN:AUTO:NOIS:TYPE Optional: if the default channel is active then set the specific InBand noise measurement type. CALC:WDM:CHAN:SEL "C_001" CALC:WDM:CHAN:NOIS:AUTO OFF Set the selected channel auto noise to disabled. CALC:WDM:CHAN:NOIS:TYPE INB Set the selected channel specific InBand noise CALC:WDM:CHAN:SEL "C_002" CALC:WDM:CHAN:NOIS:AUTO OFF Set the selected channel auto noise to disabled. CALC:WDM:CHAN:NOIS:TYPE INBN Set selected channel specific InBand noise measurement type CALC:WDM:CHAN:SEL "C_003" CALC:WDM:CHAN:NOIS:AUTO OFF Set the selected channel auto noise to disabled. Command Sequence Comments
Acquisition Sequence (WDM) on page 603.
OFF
INB
measurement type.