SCPI Command Reference

OSA

525

Product-Specific Commands—Description

:INITiate:CONTinuous?

Description

This query indicates if the trigger system is

continuously initiated or not.

At *RST, this value is set to off (disabled).

Syntax

:INITiate:CONTinuous?

Parameter(s)

None

Response Syntax

Response(s)

Continuous:
The response data syntax for is

defined as a

element.

The response corresponds to the

state of the trigger system continuous cycle.

0: continuous cycle is disabled.

1: continuous cycle is enabled.

Example(s)

INIT:CONT ON
INIT:CONT? Returns 1 (trigger system

continuously initiated)

Notes

An acquisition may still be in progress even if

INIT:CONT? returns 0. The command

STAT:OPER:BIT8:COND? shall be used to test

acquisition completion.

See Also

:INITiate[:IMMediate]

:INITiate:CONTinuous

:STATus?