SCPI Command Reference

430

FTB-5240S/S-P/BP

Product-Specific Commands—Description

DEFault allows the instrument to select a value

for the parameter.

The parameter corresponds to a

valid distance in meter from peak to center of

noise region.

The

CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce[:WAVelength]? MIN and

CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce[:WAVelength]? MAX queries can be used to

determine valid distance values.

Example(s)

CALC:WDM:CHAN:DEF "CWDM_3",1410.0 NM

CALC:WDM:CHAN:SEL "CWDM_3"

CALC:WDM:CHAN:NOIS:TYPE POLY5

CALC:WDM:CHAN:NOIS:DIST:WAV 40.0 NM

CALC:WDM:CHAN:NOIS:DIST:WAV? Returns

4.000000E-008

Notes

Custom noise measurement distance is applied

only if selected noise type is POLYnomial5.

See Also

:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTan

ce[:WAVelength]?

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDT

h[:WAVelength]

:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE

:CALCulate[1..n][:WDM]:CHANnel:SELect

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:

DISTance[:WAVelength]

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:DISTance[:WAVelength]