SCPI Command Reference

522

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:INITiate:CONTinuous

Description

This command is used to select whether the

trigger system is continuously initiated or not.

The trigger system is used to control trace

acquisition.

At *RST, this value is set to off (disabled).

Syntax

:INITiate:CONTinuous

Parameter(s)

Continuous:
The program data syntax for is

defined as a element.

The special forms ON and OFF

are accepted on input for increased readability.

ON corresponds to 1 and OFF corresponds to 0.

The parameter corresponds to

the new state of the trigger system continuous

cycle.

0 or OFF: disable continuous cycle: trigger

system returns to idle.
1 or ON: enable continuous cycle.