beautypg.com

Calculate[1..n][:wdm]:channel: noise:type – EXFO FTB -5240S/5240BP Optical Spectrum Analyzer User Manual

Page 448

background image

SCPI Command Reference

442

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:TYPE

Description

This command selects noise measurement type

for WDM analysis of the selected channel.

At *RST, this value is not available.

At CALCulate[1..n][:WDM]:CHANnel[:DEFine],

this value is set to IEC.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE

IEC|INBand|INBandNarrowfilter

Parameter(s)

Type:
The program data syntax for the first parameter is

defined as a

element. The allowed

DATA> elements for this parameter are:

IEC|INBand|INBandNarrowfilter.

The parameter corresponds to the newly

selected noise type.

IEC: selects IEC noise type.

INBand: selects InBand noise type.

INBandNarrowfilter: selects InBand narrow filter

noise type.

POLYnomial5: selects 5th order polynomial fit

noise type.