2 dip switches – Keri Systems BioPointe Users Manual User Manual
Page 53
![background image](/manuals/697233/53/background.png)
Configuring the BioPointe
52
6.2 DIP Switches
There are three DIP switches on the control board. The usage of each switch is as tabulated:
SW1
8-way
To set the Device ID
SW2
4-way
To set the baud rate and communication type
SW3
8-way
To set the type of serial communication (i.e.
whether RS232, 422 or 485)
The following tables show the settings for each of the three DIP switches.
Note that a "-"indicates an "OFF" position.
Table 1: SW1 (Device ID)
SW1 is used to set the Device ID of the device. It has a row of 8 DIP switches. The range of Device IDs that
can be set is 0 to 255. However, the acceptable range for field operation is 1 to 254. The Device ID is set
according to the binary format of an 8-bit byte. An "ON" position is interpreted as binary bit 0. To help you
to find the dip switching settings for a particular Device ID, use the index value associated with each bit.
Switch /
Device ID
1
2
3
4
5
6
7
8
Index Value
1
2
4
8
16
32
64
128
0
(Not
used) ON ON ON ON ON ON ON ON
1
-
ON
ON
ON
ON
ON
ON
ON
2
ON - ON
ON
ON
ON
ON
ON
3
-
-
ON
ON
ON
ON
ON
ON
4
ON
ON - ON
ON
ON
ON
ON
5
-
ON
-
ON
ON
ON
ON
ON
6
ON - - ON
ON
ON
ON
ON
7
-
-
-
ON
ON
ON
ON
ON
8
ON ON ON - ON ON ON ON
9
-
ON
ON
-
ON
ON
ON
ON
10
ON - ON - ON
ON
ON
ON
:
:
:
:
:
:
:
:
:
253
- ON - - - - - -
254
ON
-
-
-
-
-
-
-
255
(Not
used) - - - - - - - -
Examples:
Switch/
Device ID
1 2 3 4 5 6 7 8
Index
Value
1 2 4 8 16 32 64 128
Device
ID
5 = 1 + 4
Position
- ON - ON ON ON ON ON
Device
ID
6 = 2 +
4
Position
ON - - ON ON ON ON ON
Device ID 10
=
2
+
8
Position ON - ON
- ON
ON
ON
ON
Device
ID
30 =
2 +
4 +
8 +
16
Position ON - - - -
ON
ON
ON