MENU
BUTTON
MAIN MENU

MIN/MAX*
BUTTON SUB
MENU

CONDITION

LOCKOUT CONFIGURATION #1

“L1C.1=0”

Setpoint 1 change unlocked.

“L1C.1=1”

Setpoint 1 change locked out.

“L1C.2=0”

Setpoint 2 change unlocked.

“L1C.2=1”

Setpoint 2 change locked out.

“L1C.3=0”

Setpoint 3 (Alarm 1) change unlocked.

“L1C.3=1”

Setpoint 3 (Alarm 1) change locked out.

“L1C.4=0”

Setpoint #4 (Alarm 2) change unlocked.

“L1C.4=1”

Setpoint #4 (Alarm 2) change locked out.

“L1C.5=0”

Valley-value (LO RDG) display is permitted.

“L1C.5=1”

Displays microprocessor revision.

“L1C.6=0”

Peak-value (HI RDG) display is permitted.

“L1C.6=1”

Peak-value (HI RDG) display is not permitted.

“L1C.7=0”

INPUT CLASS (BRIDGE) is selected.

“L1C.7=1”

INPUT CLASS is locked out.

“L1C.8=0”

Not used in “BRIDGE” mode, must be set to “0”.

“L1 CNF”:

“L1C.8=1”

Not used in “BRIDGE” mode.

* The ‘MIN’ button allows you to sequence through L1C.1, L1C.2, L1C.3, L1C.4, L1C.5, L1C.6, L1C.7 and L1C.8.

The ‘MAX’ button allows you to select the “0” or “1” state for each “L1C” condition.

The ‘MENU’ button stores the selected values for all “L1C” condition(s) changed and advances the meter to “L2
CNF”. Do not press the ‘MENU’ button after each change within the submenu or the meter will advance to the next
menu item.

Every underlined “0” or “1” state is the factory preset value.

MENU
BUTTON
MAIN MENU

MIN/MAX*
BUTTON
SUB MENU

DESCRIPTION

LOCKOUT CONFIGURATION #2

“L2C.1=0”

RDG.CNF (scale/offset method and display features) may be chosen.

“L2C.1=1”

RDG.CNF (scale/offset method and display features) is locked out.

“L2C.2=0”

Either RDG SC (computed input-to-display scale factor) or RD.SC.OF (two data
points, which determine the reading scale/offset) may be entered.

“L2C.2=1”

Either RDG SC (computed input-to-display scale factor) or RD.SC.OF (two data
points, which determine the reading scale/offset) is locked out.

“L2C.3=0”

RDG OF (offset computed in display digits) may be entered.

“L2C.3=1”

RDG OF (offset computed in display digits) is locked out.

“L2C.4=0”

INP.CNF (meter rates, front end features, prelinearizing scale/offset) may be
chosen.

“L2C.4=1”

INP.CNF (meter rates, front end features, prelinearizing scale/offset) may not be
locked out.

“L2C.5=0”

IN.SC.OF (two data points for additional scale/offset) may be entered.

“L2C.5=1”

IN.SC.OF (two data points for additional scale/offset) may not be entered

“L2C.6=0”

DEC PT (decimal-point location) may be chosen

“L2C.6=1”

DEC PT (decimal-point location) may not be chosen.

“L2C.7=0”

CNT BY (round off of display) can be specified.

“L2 CNF”:

“L2C.7=1”

CNT BY (round off of display) cannot be specified.