beautypg.com

Maintenance – LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 82

background image

allowed to air-dry in an inverted position to prevent dust from entering

the tubes. Dry tubes should be stored with the caps on to prevent

contamination.
After a tube has been fi lled and capped, it should be held by the cap

and the outside surface should be wiped with a clean, lint-free absorbent

cloth until it is dry and smudge-free. Handling the tube only by the cap

will avoid problems from fi ngerprints. Always set the clean tube aside

on a clean surface that will not contaminate the tube. It is imperative that

the tubes and light chamber be clean and dry. The outside of the tubes

should be dried with a clean, lint-free cloth or disposable wipe before

they are placed in the meter chamber.
Tubes should be emptied and cleaned as soon as possible after reading

a sample to prevent deposition of particulates on the inside of the tubes.

When highly accurate results are required, reduce error by designating

tubes to be used only for very low turbidity and very high turbidity

testing.
Variability in the geometry of the glassware and technique is the

predominate cause of variability in results. Slight variations in wall

thickness and the diameter of the tubes may lead to slight variations in

the test results. To eliminate this error the tubes should be placed in the

chamber with the same orientation each time.
Chambers which have been scratched through excessive use should be

discarded and replaced with a new one.

MAINTENANCE

CLEANING

Clean the exterior housing with a damp, lint-free cloth. Do not allow

water to enter the light chamber or any other parts of the meter. To clean

the light chamber and optics area, point a can of compressed air into

the light chamber and blow the pressurized air into the light chamber.

Use a cotton swab dampened with Windex

®

window cleaner to gently

swab the interior of the chamber. Do not use alcohol; it will leave a thin

residue over the optics when dry.
REPAIRS

Should it be necessary to return the meter for repair or servicing,

pack the meter carefully in a suitable container with adequate packing

material. A return authorization number must be obtained from LaMotte

Company by calling 800-344-3100 (US only) or 410-778-3100, faxing

410-778-6394 or emailing [email protected]. Often a problem can

be resolved over the phone or by email. If a return of the meter is

necessary, attach a letter with the return authorization number, meter

serial number, a brief description of problem and contact information

including phone and FAX numbers to the shipping carton. This

information will enable the service department to make the required

repairs more effi ciently.
METER DISPOSAL

Waste Electrical and Electronic Equipment (WEEE)

82

This manual is related to the following products: