CHLORINE: ANALYSIS - DPD TABLET REAGENTS

Total Chlorine

The default units are ppm and the default calibration curve is for DPD

Tablet reagents. For the most accurate results, a user calibration should

be performed. The letter

(T) in the upper right corner of the display

indicates that the meter is in the tablet DPD reagent mode. To use liquid

DPD reagents, see the Set Up instructions.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure. Press

to

scroll to

Total Chlorine.

Measure Menu

Turbidity - No Blank
Turbidity - With Blank
Free Chlorine

Total Chlorine

12:00:00 001/500

3. Press

ENTER

to select

Total

Chlorine.

Total Chlorine (T)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

4. Rinse a clean tube (0290)

three times with the sample.

Fill the tube to the 10 mL line

with the sample. Cap the

tube. Dry the tube with a lint-

free cloth.

63

Chlorine Analysis