6. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to

scan the blank. The screen

will display

Blank Done for

about 1 second and then

return to the

Color menu.

Color

Scan Blank

Scan Sample

12:00:00 001/500

7. Remove the tube from the

meter. Empty the tube. Rinse

the tube with the sample. Fill

the tube to the 10 mL line with

the sample. Cap the tube.

8. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to select

Scan Sample. The

screen will display

Reading...

for about 1 second. The result

will appear on the screen

and then return to the

Color

menu. Record the result as

color.

Color

110

cu

Scan Blank

Scan Sample

12:00:00 001/500

NOTE: The meter will remember the last scanned blank reading. It is not

necessary to scan a blank each time the test is performed. To use the

previous blank reading, instead of scanning a new one, scroll to Scan

Sample and proceed. For the most accurate results, the meter should

be blanked before each test and the same tube should be used for the

blank and the reacted sample.

DILUTION PROCEDURE

Samples and standards may be diluted with distilled or deionized water.

TESTING TIPS

1. Always use a clean test tube.
2. Turbidity will interfere with the color test. Sample may be fi ltered

before testing but results will be as true color. (See What is Color?,

page 17)

3. When testing at low concentrations use the same tube for the blank

and the sample.

76

Color Analysis