beautypg.com

LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 81

background image

The display may show the following messages:

Battery Status

More choices are available and can be viewed by

scrolling up and/or down through the display.

Header

Identifi es the current menu and information on units

and reagent systems if applicable.

Footer

In the data logging mode the number of the data point

is displayed and the total number of data points in the

memory will be shown. The footer also shows current

time and battery status

NEGATIVE RESULTS

There are always small variations in readings with analytical instruments.

Often these variations can be observed by taking multiple readings of

the same sample. These variations will fall above and below an average

reading. Repeated readings on a 0.00 sample might give readings

above and below 0.00. Therefore, negative readings are possible and

expected on samples with concentrations at or near zero. This does

not mean there is a negative concentration in the sample. It means

the sample reading was less than the blank reading. Small negative

readings can indicate that the sample was at or near the detection limit.

This is a normal variation that results in a negative reading. A large

negative reading, however, is not normal and indicates a problem.

Some instruments are designed to display negative readings as zero.

In this type of instrument, if the meter displayed zero when the result

was actually a large negative number there would be no indication that

a problem existed. For this reason, the LTC3000we/wi displays negative

numbers for turbidity.

TUBES AND CHAMBERS

The LTC3000we/wi uses one type of tube (Code 0290) for all test factors.

There is no need for a special turbidity tube.
The handling of the tubes is of utmost importance. Tubes must be clean

and free from lint, fi ngerprints, dried spills and signifi cant scratches,

especially the central zone between the bottom and the sample line.
Scratches, fi ngerprints and water droplets on the tube can cause stray

light interference leading to inaccurate results when measuring turbidity.

Scratches and abrasions will affect the accuracy of the readings. Tubes

that have been scratched in the light zone through excessive use should

be discarded and replaced with new ones.
Tubes should always be washed on the inside and outside with mild

detergent prior to use to remove dirt or fi ngerprints. The tubes should be

81

This manual is related to the following products: