beautypg.com

LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 58

background image

4. Rinse a clean tube (0290)

three times with the chlorine

standard. Fill the tube to the

10 mL line with the chlorine

standard. Cap the tube. Dry

the tube with a lint-free cloth. .

5. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube

with the index arrow on the

meter. Close the lid. Press

ENTER

to scan the blank. The

screen will display

Blank

Done for about 1 second and

then return to the

Scan Free

Chlorine Menu.

Free Chlorine (L)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

6. Remove the tube from the

meter. Add 5 drops of DPD

1A Free Chlorine reagent

(P-6740) and add 5 drops

of *DPD 1B Free Chlorine

reagent (P-6741). Cap and

invert to mix. The solution will

be pink if chlorine is present.

Read within 30 seconds.

DPD

1B

DPD

1A

7. Immediately insert the tube

into the meter. Align the index

line on the tube with the index

arrow on the meter. Close the

lid.

8. Press

ENTER

to scan the

Standard. The screen will

display

Reading... for about 1

second. The result will appear

on the screen.

Free Chlorine (L)

0.99

ppm

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

58

Chlorine Calibration

This manual is related to the following products: