5. Press

ENTER

to select

Turbidity Units.

Set Turbidity Units

NTU

ASBC
EBC

12:00:00 001/500

Available units are:

NTU (Nephelometric Turbidity Units)( LTC3000we only)

FNU (Formazin Nephelometric Units) (LTC3000wi only)

ASBC (American Society of Brewing Chemists)

EBC (European Brewery Convention)

NOTE: The meter will automatically switch to the attenuation mode

above 600 NTU or FNU. In Attenuation mode, measurements will be

made with the 180° detector only, as indicated by AU (attenuation units)

or FAU (formazin anttenuation units) on the display. AU and FAU are

numberically equivalent to NTU. See page 15 for defi nitions of turbidity

units.

6. Press

or

to scroll

to the desired units.

Set Turbidity Units

NTU

ASBC
EBC

12:00:00 001/500

7. Press

ENTER

to save the

selection. The screen will

display

Storing... for about

1 second and return to the

Turbidity Options menu.

Press

EXIT

to return to a

previous menu.

Turbidity Options

Turbidity Units

Turbidity Calibration

12:00:00 001/500

22

Options/Set Up