4. Press
ENTER
to select
Color.
Color
Scan Blank
Scan Sample
12:00:00 001/500
5. Rinse a clean tube (0290)
with color-free (distilled or
deionized) water. Fill the tube
to the 10 mL line with the
color-free water. Cap the tube.
Dry the tube with a lint-free
cloth.
6. Open the meter lid. Insert the
tube into the chamber. Align
the index line on the tube with
the index arrow on the meter.
Close the lid. Press
ENTER
to
scan the blank. The screen
will display
Blank Done for
about 1 second and then
return to the
Color menu.
Color
Scan Blank
Scan Sample
12:00:00 001/500
7. Remove the tube from the
meter. Empty the tube.
Rinse the tube with the color
standard. Fill the tube to the
10 mL line with the color
standard. Cap the tube. Dry
the tube with a lint-free cloth.
8. Open the meter lid. Insert the
tube into the chamber. Align
the index line on the tube with
the index arrow on the meter.
Close the lid. Press
ENTER
to select
Scan Sample. The
screen will display
Reading...
for about 1 second. The result
will appear on the screen
and then return to the
Color
menu.
Color
110
cu
Scan Blank
Scan Sample
12:00:00 001/500
73
Color Calibration