LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual
Page 74
9. Press
to scroll to
Calibrate.
Color
110
cu
Scan Sample
Calibrate
12:00:00 001/500
10. Press
ENTER
to select to
Calibrate. A reverse font
(dark background with light
characters) will appear to
indicate that the reading can
be adjusted.
Color
110
cu
Scan Blank
Calibrate
12:00:00 001/500
11. Press
or
to
scroll to the concentration
of the standard, 100 in this
example. Note: The allowable
adjustment is ±25%.
Color
110
cu
Scan Blank
Calibrate
12:00:00 001/500
12. Press
ENTER
to select to
Calibrate. Two menu choices
will be offered, Set Calibration
and Factory Settings.
Calibrate Menu
110
cu
Set Calibration
Factory Setting
12:00:00 001/500
13. Press
ENTER
to select
Set
Calibration. and save the
calibration. Press
or
to scroll to and select
Factory Setting to revert to
the factory calibration. The
meter will momentarily display
Storing... and return to the
Color menu. The calibration
has now been saved and the
meter can be used for testing.
Color
Scan Blank
Scan Sample
12:00:00 001/500
74
Color Calibration