beautypg.com

LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 72

background image

COLOR

CALIBRATION

Color Standards

The meter has been calibrated with colored standards of known

concentrations of platinum cobalt. One unit of color is equivalent

to the color that is produced by 1 mg platinum/L in the form of the

chloroplatinate ion. A 500 cu Color Standard (60 mL, Code 6058-H) is

available from LaMotte.

Calibration Procedure

The meter should be calibrated with platinum cobalt color standards.

For the most accurate results, a user calibration should be performed

with LaMotte Color Standards. The calibration should be done with

a distilled or deionized water blank and one color standard of known

concentration. The concentration of the calibration standard should be

similar to the expected concentration of samples that will be tested.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure.

Measure Menu

Turbidity - No Blank

Turbidity - With Blank
Free Chlorine
Total Chlorine

12:00:00 001/500

3. Press

to scroll to

Color.

Measure Menu

Turbidity - No Blank
Free Chlorine
Total Chlorine

Color

12:00:00 001/500

72

Color Calibration