2. Press
ENTER
to select
Select
Measure. Press
to
scroll to
Total Chlorine.
Measure Menu
Turbidity - No Blank
Turbidity - With Blank
Free Chlorine
Total Chlorine
12:00:00 001/500
3. Press
ENTER
to select
Total
Chlorine.
Total Chlorine (L)
Scan Blank
Scan Free Chlorine
12:00:00 001/500
4. Rinse a clean tube (0290)
three times with the sample.
Fill the tube to the 10 mL line
with the sample. Cap the
tube. Dry the tube with a lint-
free cloth.
5. Open the meter lid. Insert
the tube into the chamber.
Align the index line on the
tube with the index arrow on
the meter. Close the lid. Press
ENTER
to scan the blank. The
screen will display
Reading...
for about 1 second and then
return to the
Total Chlorine
menu.
Total Chlorine (L)
Scan Free Chlorine
Scan Total Chlorine
12:00:00 001/500
69