beautypg.com

LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 62

background image

9. Remove the tube from the

meter. Add one *Chlorine

DPD #3 Instrument Grade

Tablet (6197A). Cap the tube.

Shake for 10 seconds. Invert

slowly 5 times. An increase in

color represents Combined

Chlorine.

NOTE: For wastewater samples, Standard Methods for the Examination

of Water and Wastewater recommends waiting 2 minutes for full color

development when testing total chlorine.

10. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

11. Press

ENTER

to select

Scan

Total Chlorine. The screen

will display

Reading... for

about 1 second. The result

will appear on the screen.

Record the result as Total

Chlorine.

Total Chlorine (L)

1.25

ppm

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

12. Subtract the Free Chlorine reading from the Total Chlorine reading

to obtain the concentration of Combined Chlorine.

Total Chlorine - Free Chlorine = Combined Chlorine

*Warning: Reagents marked with an * are considered to be potential

health hazards. To view or print a Material Safety Data Sheet (MSDS)

for these reagents go to www.lamotte.com. To obtain a printed copy,

contact LaMotte by e-mail, phone or fax.
NOTE: For the most accurate results, samples over 6 ppm chlorine

should be diluted with chlorine demand free water and re-tested.
NOTE: The meter will remember the last scanned blank reading. It is not

necessary to scan a blank each time the test is performed. To use the

previous blank reading, instead of scanning a new one, scroll to Scan

Sample and proceed. For the most accurate results, the meter should

be blanked before each test and the same tube should be used for the

blank and the reacted sample.

62

Chlorine Analysis

This manual is related to the following products: