beautypg.com

LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 61

background image

4. Rinse a clean tube (0290)

three times with the sample.

Fill the tube to the 10 mL line

with the sample. Cap the

tube. Dry the tube with a lint-

free cloth.

5. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to

scan the blank. The screen

will display

Blank Done for

about 1 second and then

return to the

Free Chlorine

menu.

Free Chlorine (T)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

6. Remove the tube from the

meter. Add one *Chlorine

DPD #1 Instrument Grade

Tablet (6903A). Cap the tube.

Shake for 10 seconds. Invert

slowly 5 times. The solution

will be pink if Free Chlorine is

present.

7. Immediately insert the tube

into the meter. Align the index

line on the tube with the index

arrow on the meter. Close the

lid.

8. Press

ENTER

to scan the

sample. The screen will

display

Reading... for about 1

second. The result will appear

on the screen. Record the

result as Free Chlorine.

Free Chlorine (T)

0.99

ppm

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

61

Chlorine Analysis

This manual is related to the following products: