LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 64

background image

5. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid. Press

ENTER

to

scan the blank. The screen

will display

Blank Done for

about 1 second and then

return to the

Total Chlorine

menu.

Total Chlorine (T)

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

6. Remove the tube from the

meter. Add one *Chlorine

DPD #1 Instrument Grade

Tablet (6903A). Cap the tube.

Shake for 10 seconds. Invert

slowly 5 times. The solution

will be pink if Free Chlorine is

present.

7. Add one *Chlorine DPD #3

Instrument Grade Tablet

(6197). Cap the tube. Shake

for 10 seconds. Invert slowly

5 times. The solution will

be pink if Free Chlorine is

present.

NOTE: For wastewater samples, Standard Methods for the Examination

of Water and Wastewater recommends waiting 2 minutes for full color

development when testing total chlorine.

8. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

64

Chlorine Analysis