beautypg.com

LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 51

background image

simplicity the LTC3000we/wi uses the term turbidity-free water and the

value of 0.00 NTU.
PREPARATION OF TURBIDITY-FREE WATER

A 0 NTU/FNU Standard (Code 1480) is included with the meter. An

accessory package (Code 4185) is available for preparing turbidity-free

water for blanking the meter and dilution of high turbidity samples.
The preparation of turbidity-free water requires careful technique.

Introduction of foreign matter will affect the turbidity reading. A fi ltering

device with a special membrane fi lter is used to prepare turbidity-

free water. The fi lter, fi lter holder and syringe must be conditioned

by forcing at least two syringes full of deionized water through the

fi ltering apparatus to remove foreign matter. The fi rst and second

rinses should be discarded. Turbidity-free water as prepared with the

following procedure may be stored in the dark at room temperature in a

clean glass bottle with a screw cap and used as required. The storage

container should be rinsed thoroughly with fi ltered deionized water

before fi lling. The water should be periodically inspected for foreign

matter in bright light.

1. Remove the plunger from

the syringe (0943). Attach

the fi lter to the bottom of the

syringe.

2. Pour approximately 50 mL

of deionized water into the

barrel of the syringe. Insert

the plunger. Exert pressure

on the plunger to slowly force

the water through the fi lter.

Collect water in the clean

storage container. Rinse walls

of the container then discard

this rinse water.

51

Turbidity Analysis

This manual is related to the following products: