beautypg.com

General operating information – LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual

Page 78

background image

STRAY LIGHT

The accuracy of readings on the LTC3000we/wi should not be affected

by stray light. Make sure that the sample compartment lid is always fully

closed when taking readings. The backlight will interfere with turbidity

readings. The meter will temporarily disable the backlight while turbidity

measurements are being taken.

GENERAL OPERATING INFORMATION

OVERVIEW

The LTC3000we/wi is a portable, microprocessor controlled, direct

reading nephelometer. Turbidity is measured directly by either EPA

Method 180.1 or ISO Method 7027. It has a graphical liquid crystal

display and 6 button keypad. These allow the user to select options from

the menu driven software, to directly read test results or to review stored

results of previous tests in the data logger. The menus can be displayed

in seven different languages.
The LTC3000we/wi uses a state of the art, multi-detector optical

confi guration that assures long term stability of calibrations, high

precision and accuracy and low detection limits. All readings are

determined by sophisticated digital signal processing algorithms,

minimizing fl uctuations in readings and enabling rapid, repeatable

measurements. The microprocessor and optics enable a dynamic range

and auto-ranging over several ranges. Energy effi cient LED light sources

are used for ISO turbidity. EPA turbidity uses a tungsten fi lament light

source that meets or exceeds EPA specifi cations and is designed for a

uniform light spot image and stable output.
A USB computer/wall charger or Lithium battery powers the

LTC3000we/wi.
A USB port on the back of the meter allows an interface of the meter

with a Windows-based computer for real-time data acquisition and data

storage using a PC. The LTC3000we/wi may be interfaced with any

Windows-based computer by using the LaMotte SMARTLink 3 Program.
GENERAL OPERATING INFORMATION

The operation of the LTC3000we/wi is controlled by the menu driven

software and user interface. A menu is a list of choices. This allows a

selection of various tasks for the LTC3000we/wi to perform, such as scan

blank and scan sample. The keypad is used to make menu selections

that are viewed on the display.

78