*WARNING: Reagents marked with an * are considered to be potential

health hazards. To view or print a Material Safety Data Sheet (MSDS)

for these reagents go to www.lamotte.com. To obtain a printed copy,

contact LaMotte by e-mail, phone or fax.
NOTE: The meter will remember the last scanned blank reading. It is not

necessary to scan a blank each time the test is performed. To use the

previous blank reading, instead of scanning a new one, scroll to Scan

Sample and proceed. For the most accurate results, the meter should

be blanked before each test and the same tube should be used for the

blank and the reacted sample.

CHLORINE: ANALYSIS - DPD TABLET REAGENTS

Free Chlorine, Combined Chlorine and Total Chlorine

The default units are ppm and the default calibration curve is for DPD

Tablet reagents. For the most accurate results, a user calibration should

be performed. The letter (

T) in the menu bar indicates that the meter

is in the tablet mode. To use liquid DPD reagents, see the Set Up

instructions.

1. Press and briefl y hold

to turn the meter on. The

LaMotte logo screen will

appear for about 3 seconds

and the

Main Menu will

appear.

Main Menu

Measure

Data Logging
Options
Run PC Link

12:00:00 001/500

2. Press

ENTER

to select

Measure. Press

to scroll

to

Free Chlorine.

Measure Menu

Turbidity - No Blank
Turbidity - With Blank

Free Chlorine

Total Chlorine

12:00:00 001/500

3. Press

ENTER

to select

Free

Chlorine.

Free Chlorine (T)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

60

Chlorine Analysis