LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual
Page 32
SETTING THE BACKLIGHT TIME
The backlight illuminates the display for enhanced viewing. If Button
Control is chosen the backlight button on the key pad will act as an
on/off switch and the backlight will remain on or off when the meter is
being used. When one of the other settings – 10, 20 or 30 seconds –
is chosen, the display will be illuminated for the specifi ed amount of
time after any button is pressed. As a precaution, the backlight will not
illuminate during turbidity measurements to avoid interference from stray
light.
NOTE: The backlight feature uses a signifi cant amount of power. The
longer the backlight is on, the more frequently the battery will have to be
charged if the USB/Wall Charger is not being used.
1. Press and briefl y hold
to turn the meter on. The
LaMotte logo screen will
appear for about 3 seconds
and the
Main Menu will
appear.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
2. Press
to scroll to
Options.
Main Menu
Measure
Data Logging
Options
Run PC Link
12:00:00 001/500
3. Press
ENTER
to select
Options.
Options Menu
Averaging
Turbidity Options
Chlorine Options
Set Clock
12:00:00 001/500
32
Options/Set Up