8. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

Press

ENTER

to scan the blank.

The screen will display

Blank

Done for about 1 second and

then return to the

Turbidity - With

Blank menu.

Turbidity WB (F)

Scan Blank

Scan Sample

12:00:00 001/500

9. Rinse a clean tube (0290), or

the same tube, three times

with the sample.

l

10. Fill the tube to the fi ll line

with the standard. Pour the

standard down the inside of

the tube to avoid creating

bubbles. Cap the tube.

11. Wipe the tube thoroughly with

a lint-free cloth.

49

Turbidity Analysis