9. Press

to scroll to

Calibrate.

Free Chlorine (L)

0.99

ppm

Scan Total Chlorine

Calibrate

12:00:00 001/500

10. Press

ENTER

to select

Calibrate. A reverse font

(dark background with light

characters) will appear to

indicate that the reading can

be adjusted.

Free Chlorine (L)

0.99

ppm

Scan Total Chlorine

Calibrate

12:00:00 001/500

11. Press

or

to

scroll to the concentration

of the standard, 1.00 in this

example.

Note: The allowable

adjustment is ±25%.

Free Chlorine (L)

1.00

ppm

Scan Total Chlorine

Calibrate

12:00:00 001/500

12. Press

ENTER

to select

Calibrate. Two menu choices

will be offered, Set Calibration

and Factory Settings.

Calibrate Menu

1.00

ppm

Set Calibration

Factory Setting

12:00:00 001/500

13. Press

ENTER

to select

Set

Calibration and save the

calibration. Press

or

to scroll to and select

Factory Setting to revert to

the factory calibration. The

meter will momentarily display

Storing... and return to the

Free Chlorine menu. The

calibration has now been

saved and the meter can be

used for testing.

Free Chlorine (L)

Scan Blank

Scan Free Chlorine

12:00:00 001/500

59

Chlorine Calibration