LaMotte LTC3000wi Turbidity & Chlorine Lab Meter User Manual
Page 24
![background image](/manuals/596704/24/background.png)
24
6. Press
ENTER
to select
Turbidity Calibration.
Turbidity Calibration
Formazin
Japan Standard
12:00:00 001/500
7. Scroll to the desired
calibration option. Select
a calibration option based
on the composition of the
standards that will be used to
calibrate the meter.
Turbidity Calibration
Formazin
Japan Standard
12:00:00 001/500
NOTE: Stablcal® standards below 50 NTU should not be used to
calibrate the LTC3000we/wi. The diluent has a different refractive index
than traditional formazin standards and will affect the results.
8. Press
ENTER
to save the
selection. The screen will
display
Storing... for about
1 second and return to the
Turbidity Options menu.
Press
EXIT
to return to a
previous menu.
Turbidity Options
Turbidity Units
Turbidity Calibration
12:00:00 001/500
Options/Set Up