6. Remove the tube from the

meter. Add 5 drops of DPD

1A Free Chlorine reagent

(P-6740) and add 5 drops of

DPD 1B Free Chlorine reagent

(P-6741). Cap and invert to

mix. The solution will be pink

if chlorine is present.

DPD

1B

DPD

1A

7. Add 5 drops of *DPD 3 Total

Chlorine reagent (P-6741).

Cap and invert to mix. An

increase in color represents

Combined Chlorine.

DPD

3

NOTE: For wastewater samples, Standard Methods for the Examination

of Water and Wastewater recommends waiting 2 minutes for full color

development when testing total chlorine.

8. Open the meter lid. Insert the

tube into the chamber. Align

the index line on the tube with

the index arrow on the meter.

Close the lid.

9. Press

ENTER

to select

Scan

Total Chlorine. The screen

will display

Reading... for

about 1 second and then

return to the

Total Chlorine

menu. Record the result as

Total Chlorine.

Total Chlorine (L)

1.25

ppm

Scan Free Chlorine

Scan Total Chlorine

12:00:00 001/500

*Warning: Reagents marked with an * are considered to be potential

health hazards. To view or print a Material Safety Data Sheet (MSDS)

70

Chlorine Analysis