PROBLEMS/SERVICE

TRANSMIG 250i

Manual 0-5187

5-5

PROBLEMS AND ROUTINE SERVICE

Power Source Calibration

A. Schedule

Output testing of all CIGWELD Inverter Power Sources and applicable accessories shall be conducted at regular

intervals to ensure they fall within specified levels. Calibration intervals shall be as outlined below -

1. For transportable equipment, at least once every 3 months; and
2. For fixed equipment, at least once every 12 months.

If equipment is to be used in a hazardous location or environments with a high risk of electrocution as outlined

in AS 1674.2 - 2007, then the above tests should be carried out prior to entering this location.

B. Calibration Requirements

Where applicable, the tests outlined in Table 5-4 below shall be conducted by an accredited CIGWELD service

agent.

Testing Requirements

Output current (A) to be checked to ensure it falls within applicable CIGWELD power source specifications
Output Voltage (V) to be checked to ensure it falls within applicable CIGWELD power source specifications
Motor Speed (RPM) of wire drive motors to be checked to ensure it falls within required CIGWELD power

source / wire feeder specifications
Accuracy of digital meters to be checked to ensure it falls within applicable CIGWELD power source specifi-

cations

Table 5-4: Calibration Parameters

Periodic calibration of other parameters such as timing functions are not required unless a specific fault has

been identified.

C. Calibration Equipment

All equipment used for Power Source calibration shall be in proper working condition and be suitable for con-

ducting the measurement in question. Only test equipment with valid calibration certificates (NATA certified

laboratories) shall be utilized.