beautypg.com

Waters MassPREP Phosphopeptide Standard User Manual

Care and use manual, Massprep phosphopeptide standard – enolase, I. introduction

background image

[ Care and Use ManUal ]

MassPREP Phosphopeptide Standard – Enolase

I. IntroductIon

As a tool for screening sample preparation and sample enrichment methods
for phosphopeptides, Waters offers the MassPREP

Phosphopeptide Standard

– Enolase. Commonly used phosphorylated samples are often beta-casein
digests or standards which contain only two phosphopeptides, both modified
at serine. The enolase phosphopeptide standard contains four modified enolase
peptides, singly phosphorylated at either serine, threonine, or tyrosine or dou-
bly phosphorylated at serine. The phosphopeptides are packaged as lyophilized
solids in glass vials. The vials are vacuum-sealed in foil pouches to ensure
protection from light and air.

II. recommended usage

Each vial contains nmol each of four synthetic enolase phosphopeptides
(T8 (P, tyrosine), T9 (P, serine), T43 (P, threonine), T43 (2P, serine)).
This standard was prepared by purifying bulk synthetic peptide from GenScript
Corp. (Piscataway, NJ). Using this standard, the user can test phosphopeptide
detection without competing components as well as add these phosphopep-
tides to any desired mixture. The standard is intended for use in optimizing
phosphopeptide detection in liquid chromatography (UV or MS detection) and
matrix-assisted laser desorption/ionization (MALDI).

Table 1: Synthetic Enolase Phosphopeptides

III. sample preparatIon

The following methods are provided as a general guideline. For optimum
performance, use of the highest purity commercially available solvents is
recommended (HPLC grade or better).

a. Procedure

. Add 00 μL of water or other preferred solvent to the glass vial

containing the lyophilized phosphopeptide standard to make a 0
pmol/µL solution of each phosphopeptide.

2. Vortex the vial for about 30 seconds to ensure that the solid is

completely dissolved.

3. Make the desired number of dilutions in the preferred solvent.

4. Prior to MALDI, use a standard sample (peptide mix or protein

digest) to adjust instrument settings for optimum resolution and
sensitivity for the mass range between m/z 800 and 2000.

a. For MALDI, make a solution of 20 mg/mL 2,5-DHB matrix in

4: (v/v) ACN:water with 0.% TFA.

b. Apply μL (or less) of the phosphopeptide standard

(~ pmol/µL) onto a clean MALDI target. Add an equal
volume of matrix solution to the sample on the target plate.
Dry at room temperature before submitting target for MALDI
analysis.

5. Prior to HPLC, use enolase digest to adjust instrument settings

for optimum resolution and sensitivity. Adjust parameters so that
peak shape and detector sensitivity are optimized.

a. For HPLC, transfer desired amount of solution to appropriate

sample vial and subject to LC-UV and/or LC-MS analysis.

i. For separations on ~2 mm diameter columns, the recom-

mended injection volume is 5 µL or more of a solution 0
pmol/µL or more.

ii. For separations on columns less than 200 um in diameter,

the recommended injection volume is µL of a solution
200 fmol/µL or more.

massprep phosphopeptIde standard – enolase

Peptide

Sequence

Average MW

(g/mol)

Isotopic mass,

[M+H]

+

Isotopic mass,

[M+2H]2

+

T8 P or T8p

NVPLpYK

82.858

83.392

407.995

T9 P or T9p

HLADLpSK

862.875

863.4028

432.2053

T43 P or T43p

VNQIGpTLSESIK

368.444

368.6776

684.8428

T43 2P or T43pp VNQIGTLpSEpSIK

448.424

448.6439

724.8259