[ Care and Use ManUal ]

nanoACQUITY UPLC Column Care and Use

7

Replace the Emitter

Place the pin plug at the back side of the union. This will mark the

center of the union (right)

Thread the replacement emitter (back end first) through the front of the

sprayer and into the union

Tighten the PEEK nut around the union carefully (do not over-tighten)

Remove the pin plug and replace it with the transfer line from the

nanoACQUITY column (again, do not over-tighten the fitting).

About poorly cut fittings

DO NOT cut or modify the emitter or column tubing in any way.

All tubing and columns supplied by Waters is pre-cut and polished.

Improperly cut tubing can lead to "shedding" and introduction of
fused silica particles downstream of nanoACQUITY 75, 100, or
150um I.D. column. This consequently can cause undesired system
backpressure increases.

To reduce the occurrence of clogging

Make sure the union between the column and emitter is clean

Do not over tighten the fittings on the union (finger tight only)

Make sure the UPLC solvents are clean

Keep a flow on the tip even when the instrument is not in use (as low as

200 nL/min)

Try using a TaperTip instead of a Pulled Tip Emitter if clogging persists

TaperTip Emitter

A TaperTip (P/N 186003932) is offered as an alternative to a
PicoTip and is the default emitter in the universal sprayer kit. (July,
2007). The TaperTip has an ID of 20 µm but unlike the PicoTip the ID
remains unchanged thought out the emitter. This allows the emitter
to be implemented for higher flow rates. Furthermore, the TaperTip
may experience a lower frequency of particle clogging.

Note: Compared to the PicoTip Emitter, Waters TaperTip Emitter does
not produce as stable a spray at 250-400 nL/min, which is the flow
rate range most customers use with 75 µm or 100 µm ID columns.
As indicated above, use of the TaperTip Emitter is the suggested
when a more robust device is preferred.

Column ID

Optimal Flow Rate (nL/min)

Emitter Type

75 µm

300

PicoTip

100 µm

400-500

PicoTip / TaperTip

150 µm

1000

TaperTip

300 µm

4000

ESI Source required