[ Care and Use ManUal ]

Carbamate Analysis Column

5

Table 1: Troubleshooting the Carbamate Analysis Column

b. Shutdown and Storage

Between Analyses

During the course of a working day, between analyses, continue
to pump the initial mobile phase mixture through the column. This
will maintain the equilibrium in the column necessary for good
retention time reproducibility. If a few hours will pass before the
next injection, the flow rate may be slowed down in the interim to a
few tenths of a mL/min to conserve solvent.

Overnight

When shutting down overnight or over a weekend, first flush the
column with 15-30 mL of 50/50 v/v acetonitrile/methanol. Then
turn the flow rate to zero mL/min and leave the column connected
in the system. The oven temperature may be maintained at 30 °C,
if desired.

Long-term Storage (more than 72 hours)

Flush the column with 15-30 mL of 50/ v/v acetonitrile/methanol.
Then turn the pump off. Allow the column to cool to ambient
temperature. Disconnect the inlet and outlet tubes from the column
and join them with a union. Install the end plugs in the column inlet
and outlet fittings. Tighten the end plugs firmly in place with 5/16”
open end wrench.

CAUTION: Do not overtighten – Overtightening will damage the
column fittings. Allowing columns to dry out may result in poor
chromatographic performance. Return the column to its box for
storage

VI. ColuMn and supplIes orderIng InforMatIon

Symptom

Conditions

Corrective Action

Excess pressure
buildup

Filters plugged with
particulates.

Replace filter element or clean
in an ultrasonic bath. Always
filter solvents and samples

Sample precipitates on
column (sample not
soluble in mobile
phase).

Slowly purge with a solvent
appropriate to dissolve the
precipitate.

Clogged tubing.

Replace tubing.

Loss of resolution,
broad peaks, or
low plate counts

Mass overload

Dilute sample and run it again.

Incorrect tubing size
(0.040” or 0.020”
i.d.)

Install correct 0.009” ID stain-
less steel tubing to column
inlet & outlet.

Contaminated column.

Slowly flush with a 50-100
mL of 50/50 v/v acetonitrile/
MeOH; then equilibrate with
initial mobile phase and run
sample again.

Insufficient equilibrium Continue equilibration
Filters partially plugged Replace or clean (inlet and

outlet) the Filter Retainer Disc
and the Filter.

Failing Injector

Repair Injector.

Item

Part Number

Carbamate Analysis Column, 3.9x150mm

WAT035577

Filter Retainer

WAT088084

Filter Retainer Disc

WAT089567

In-Line Pre column Filter

WAT084560

Solvent Clarification Kit with pump

WAT085113

Solvent Clarification Kit without pump

WAT085124

Aqueous Replacement Filters (HATF 04700),
pkg. of 100

WAT085147

Organic Replacement Filters (FHUP 04700),
pkg. of 100

WAT085118

Aqueous Sample Clarification Kit

WAT026865

Guard-Pak Pre column Module: Kit

WAT080040

Nova-Pak

®

C

18

Guard-Pak Cartridges (10/Pkg.)

WAT015220

Sep-Pak C

18

Cartridges (50/Box)

WAT051910

Sep-Pak C

18

Plus Cartridges (24/Box)

WAT011191

Sep-Pak C

18

Plus Cartridges (96/Box)

WAT015402