[ Care and Use ManUal ]

Ultrahydrogel Columns

2

II. InstallatIon

a. Solvents

Ultrahydrogel columns are packed in distilled water for shipping.
Table 1 shows the maximum solvent replace¬ment flow rate for each
Ultrahydrogel column. Perform solvent changeover at a lower flow
rate than that shown in Table 2 for analysis.

CAUTION: ULTRAHYDROGEL COLUMNS ARE FLOW (PRESSURE) SENSI-
TIVE AND EXTREME CARE MUST BE TAKEN NOT TO EXCEED MAXIMUM
FLOW RATE ON START-UP.

b. Column

Table 1. Flow Rates for Replacing Solvents

NOTE: CHANGE SOLVENTS AS SELDOM AS POSSIBLE. FREQUENT
SOLVENT EXCHANGES ACCELERATE THE DEGRADATION OF COLUMN
EFFICIENCY.

c. Organic Solvents

Ultrahydrogel columns are physically and chemically stable in
ordinary water-soluble organic solvent mix¬tures. Aqueous solutions
of methanol, ethanol, acetonitrile, formic acid, and dimethylsulfoxide
also provide dependable separations. Organic solvent concentration
should not exceed 20%. It is possible however, to use organic sol-
vents up to 50% concentration by gradual introduction via gradient
method.

d. Aqueous Salt and Buffer Solution

Although some non-ionic samples can be analyzed in distilled water,
most water soluble polymers require aqueous salt or buffer solutions.

Some typical aqueous salt solutions:

Sodium sulfate

Sodium acetate

Sodium dihydrogenphosphate

Ammonium acetate

Ammonium formate

Compatible buffer solutions:

Phosphate

Tris phosphate

Tris hydrochloric acid

Citrate

Tris acetate

Acetate

e. Flow Rate

For best resolution and maximum column life the flow rate should
not exceed the maximum flow rate listed below. Use Table 2 below to
determine recommended flow rates and maximum pressures.

Table 2. Recommended Flow Rates

Column

Column Size

Max. Flow Rate During

Solvent Change

Ultrahydrogel (all pore sizes)

7.8 mm i.d. x 30 cm

0.3 ml/min

Ultrahydrogel DP

7.8 mm i.d. x 30 cm

0.3 ml/min

Ultrahydrogel DNA

7.8 mm i.d. x 30 cm

0.3 ml/min

Ultrahydrogel Linear

7.8 mm i.d. x 30 cm

0.3 ml/min

Column

Typical Operating Flow

Rate (mL/min)

Maximum Flow Rate

(mL/min)

Ultrahydrogel 120

0.5-0.8

1.0

Ultrahydrogel 250

0.5-0.8

1.0

Ultrahydrogel 500

0.3-0.6

1.0

Ultrahydrogel 1000

0.3-0.6

1.0

Ultrahydrogel 2000

0.3-0.6

1.0

Ultrahydrogel Linear

0.3-0.6

1.0

Ultrahydrogel DP

0.5-0.8

1.0