Care and use manual – Waters AccQ-Tag Amino Acid Analysis Column User Manual
Page 4

[ CARE AND USE MANUAL ]
[ CARE AND USE MANUAL ]
4
2. If the column has been stored in 100% acetonitrile, condition
the column with acetonitrile/water (60%/40%) at 1 mL/min for
5 minutes before equilibrating in Eluent A.
3. After equilibrating the column, run a blank gradient.
IV. CARE AND MAINTENANCE
a. Storing the Column
For periods of three days or less, store the column in acetonitrile/water
(60%/40%). If the column will not be used for more than 72 hours,
store it in 100% acetonitrile.
Do not allow the column to dry out or to freeze.
Symptom
Conditions
Corrective action
Excess pressure
buildup
In-line precolumn filter plugged. Check for
injector or pump seal shedding.
Replace filter.
Column frit plugged
Replace frit.
Clogged tubing
Unclog or replace tubing.
Failing injector
Repair the injector.
Column contaminated
Flush column with 100% acetonitrile at 0.5 ml/min for 30 minutes.
Loss of resolution
broad peaks,
low plate counts
Insufficient equilibration
Continue equilibration.
Column contaminated
Flush column with 100% acetonitrile at 0.5 ml/min for 30 minutes.
Incorrect diameter tubing
Install 0.005 inch PEEK
™
tubing.
Table 2. Column Problems and Solutions