Care and use manual – Waters AccQ-Tag Amino Acid Analysis Column User Manual

Page 4

background image

[ CARE AND USE MANUAL ]

[ CARE AND USE MANUAL ]

4

2. If the column has been stored in 100% acetonitrile, condition

the column with acetonitrile/water (60%/40%) at 1 mL/min for

5 minutes before equilibrating in Eluent A.

3. After equilibrating the column, run a blank gradient.

IV. CARE AND MAINTENANCE

a. Storing the Column

For periods of three days or less, store the column in acetonitrile/water

(60%/40%). If the column will not be used for more than 72 hours,

store it in 100% acetonitrile.

Do not allow the column to dry out or to freeze.

Symptom

Conditions

Corrective action

Excess pressure
buildup

In-line precolumn filter plugged. Check for
injector or pump seal shedding.

Replace filter.

Column frit plugged

Replace frit.

Clogged tubing

Unclog or replace tubing.

Failing injector

Repair the injector.

Column contaminated

Flush column with 100% acetonitrile at 0.5 ml/min for 30 minutes.

Loss of resolution
broad peaks,
low plate counts

Insufficient equilibration

Continue equilibration.

Column contaminated

Flush column with 100% acetonitrile at 0.5 ml/min for 30 minutes.

Incorrect diameter tubing

Install 0.005 inch PEEK

tubing.

Table 2. Column Problems and Solutions