beautypg.com

Care and use manual – Waters AccQ-Tag Amino Acid Analysis Column User Manual

Page 2

background image

[ CARE AND USE MANUAL ]

[ CARE AND USE MANUAL ]

2

II. INSTALLING THE COLUMN

a. Preparing the System

This procedure requires a zero dead-volume union. Before installing

the column in the flow path:

1. Directly connect the injector to the detector by replacing the old

column with a zero dead-volume union.

2. Flush the HPLC system with acetonitrile/water (60%/40%) at a

flow rate of 1.0 ml/min.

3. Remove the union.

4. Install the column in the column heater (see Section 2.2).

b. Installing the Columns

To install the column:

1. Remove the end plugs from the column and save them for use when

you store the column.

2. Screw the inlet and outlet fittings into the column until finger-

tight. An arrow on the label indicates the direction of flow.

If you are using a stainless steel system, use two 5/16 inch wrenches

to tighten the fittings 1/4 to 1/2 turn. Do not over tighten. Over tight-

ening damages the connection.

Preparing a new PEEK

connection

Prepare a new tubing-ferrule connection (Figure 2-1) if the existing

connection is damaged or worn.

To prepare a new PEEK

tubing-ferrule connection:

1. Use a single-edge razor to make a straight, square cut in front of

the compression screw on the worn fitting.

2. Slide a compression screw and washer over the end of the

cut

tubing.

Properly bottom the tubing in the fitting seat. Dead volume results in

sample bandspreading.

To prepare a new stainless steel tubing-ferrule connection:

1. Use a file with a cutting edge or a tube cutter to scribe the

circumference of the steel tubing at the desired break point.

2. Grasp the tubing on both sides of the scribe mark with cloth-

covered pliers (to prevent marring the tube surface), and gently

work the tube back and forth until it separates.

3. Ensure that the tubing end is straight, open, and free of burrs.

4. Slide the compression fitting, followed by the ferrule (large end of

the toper first) over the tube.

5. Properly bottom the tubing in the fitting seat. Dead volume results

in sample bandspreading.

Figure 1. Ferrule and Compression Screw Assembly

III. USING THE COLUMN

a. Chromatography Guidelines

Before running the first amino acid analysis on the new column,

perform the efficiency test in Section IV. b. Efficiency Testing.

General Considerations

The following operating guidelines will help you obtain the best

performance from your AccQ•Tag

column:

• Do not exceed an operating pressure of 26 MPa /260 atm or 4000 psi).

• Filter all eluents with a 0.22 pm membrane filter. Never use turbid

or cloudy eluents.

• Protect the column from vibration and mechanical shock.

Ferrule 0.009-inch I.D. Tubing (0.25 mm)

0.130-inch gap (3.3 mm)