beautypg.com

Care and use manual – Waters µStyragel Column User Manual

Page 6

background image

[ Care and Use ManUal ]

µStyragel Column

6

b. Storing the Column

If you will be using the column again in less than 24 hours, special
storage procedures are unnecessary. However, be sure that the
columns never dry out.

For longer storage periods, return the column to its box with the end
plugs firmly in place for storage. Do not leave a column at elevated
temperatures without solvent flow. Refer to Table 2 for information
about solvent compatibility for your column.

c. Efficiency Testing

Waters columns are tested for adherence to our specifications. Slight
variations may occur depending on:

The type and condition of equipment used

The nature of the sample

Instrument settings

Waters recommends that you perform your own efficiency tests. To
test both your system and each column in the system, perform both of
these tests:

• Band spreading instrument test

• Column efficiency test

These 2 tests are described below on the following page.

When to Perform Efficiency Tests

Perform each time you add columns or otherwise change your effi-
ciency tests system. Then, as you use the system, conduct efficiency
tests on a regular schedule.

If problems occur during normal operation of the column, repeat the
conditions for the initial efficiency tests and compare the results.
Differences in the results may help you to identify the source of the
problem.

Monitor band spreading due to system components frequently by
performing a column efficiency test without the column in line. Be
sure to resolve the cause of excessive band spreading before attach-
ing columns to the system.

Problem

Cause

Corrective Action

Prevention

Buildup in system operating
pressure

Inlet filter insert (from the first column
only) plugged with particulates. Check
for injector and pump seal shedding.

Replace the filter insert

Use a Styragel Guard Column
or fit an in-line filter between
the pump and the first column.

Clogged tubing

Replace the tubing

Filter the sample.

Loss of resolution, broad peaks,
low plate counts

Filter inserts partially blocked

Replace or clean inlet and/
or outlet filter inserts

Always filter samples and
solvents.

Failing injector

Repair the injector

Filter solvents and use an
in-line filter.

Insufficient equilibration

Continue equilibration

Always equilibrate system until
you have a stable baseline.

Column damaged

Replace column

Use care in changing solvents
and flow rates.

Table 5: Column Troubleshooting

V. CaRe anD maIntenanCe

a. Troubleshooting

Table 5 describes specific problems, corrective actions, and preventive measures.